There is no degradation in performance or loss of functionality following
the application of electrostatic discharges.
The approved configuration for reference orientation as viewed by the
operator is with the instrument in a horizontal position facing the noise
source where the receiver of the electromagnetic field replaces the
noise source.
All settings and orientations of the instrument have similar radio-
frequency emissions. These radio-frequency emissions are within the
specified limits of this standard.
All modes of operation and connecting device configurations have
similar immunity to power and radio frequency fields.
Pulse Range:
>53 dB
Measurement Ranges:
Display Range (dB) Primary RMS range
at 1 kHz
Peak
30 - 100 41 - 86 63 - 103
50 - 120 55 - 106 83 - 123
70 - 140 75 - 125
Defined from 10 dB
above noise floor to a
crest factor limit of 5
103 - 143
Weightings:
Frequency: A, C and linear 20Hz - 20kHz,
Time: Fast, Slow, Impulse, Peak,
Amplitude: 3, 4, 5, 6 dB exchange rates.
Clock:
Date and time can be set from the keypad.
Date and time can be set to PC time via dB10 and dB12.
Dose Measurement Timer: (CEL-360 Series only)
Use with pre-programmed delayed start and stop times, or with fixed
measurement durations can be selected from the keypad.
(Start and Stop times must be programmed via dB10 and dB12.)
Dose Measurement Start/Stop Delay Timer: (CEL-360 Series only)
Delay intervals can be set only via dB10 and dB12.
Settings can be specified for up to 16 sets of start and stop times, up to
30 days ahead.
The clock indicates the time remaining before the next run starts.
Dose Measurement Elapsed Time Timer: (CEL-360 Series only)
Measurement elapsed times (run durations) can be selected from the
keypad. The following options are available:
Disabled,
5, 10, 15, 30 minutes,
1, 2, 4, 8, 10, 12, 24 hours.
Under Range Indication:
1 dB below the bottom of the selected range.