May 2003
CEL-310 DOSE BADGE
&
CEL-315 READER UNIT
Users Handbook
060349/HB-01
CASELLA CEL
Regent House
Wolseley Road
Kempston, Bedford
MK42 7JY U.K.
Phone: +44 (0)1234 844 100
Fax: +44 (0)1234 841 490
E-Mail: info@casellacel.com
Web: www.casellacel.com
CASELLA USA
17 Old Nashua Road #15
Amherst
NH 03031
U.S.A.
Toll Free: +1 800 366 2966
Fax: +1 603 672 8053
e-mail: info@casellausa.com
Web: www.casellausa.com
Zobrazit stránku 0
1 2 3 4 5 6 ... 44 45

Shrnutí obsahu

Strany 1 - CEL-315 READER UNIT

May 2003CEL-310 DOSE BADGE&CEL-315 READER UNITUsers Handbook060349/HB-01CASELLA CELRegent HouseWolseley RoadKempston, BedfordMK42 7JY U.K.Phone: +

Strany 2 - Caution !

2. Insert usable batteries into the Reader Unit (Section 2.4).3. Use the Reader menu to check that the configuration set forthe Dose Badge is suitable

Strany 3 - Contents

1. Make sure the Dose Badge is in Stop Mode.2. Carefully screw it on to the charging stud of a CEL-6807 Charger.The five badge CEL-6807/5 charger is s

Strany 4

4. To ensure optimum operation, leave the Dose Badge on tricklecharge when not in use.2.4 Insert Batteries into the Reader UnitThe Reader Unit require

Strany 5 - 1. INTRODUCTION

The display backlight will be lit when any key is pressed and willautomatically switch off after 10 seconds to preserve batterypower.2. If the display

Strany 6 - 1.1 The System

10. Check the setting for criterion time (i.e. duration of a nominalworking day) by using to select Criterion Time.The criterion time is indicated in

Strany 7 - 1.2 Layout & Controls

3. CONFIGURE THE READER UNITBefore taking measurements with the Dose Badge system, followSection 2.5 to check that the configuration selected meets th

Strany 8 - Introduction

5. Use and to change entries to show the required time anddate.6. Once the time and date entries are correct, press again tosave the clock setting and

Strany 9 - 2.2 Quick Start

of L or TWA, %Noise Dose and Estimated %Dose. The ability toconfigure the Criterion Time enables Dose Badge measurements to bemade for a particular le

Strany 10 - Preparation

3. Use and to select the required setting (see Table 1).4. Press again to save the setting and return to the Main Menu.3.4 Configure Criterion LevelTh

Strany 11

3.5 Configure Dose Badge SettingsSettings to be used for some International Standards are given in Table 1.PLEASE NOTE that changes to the Dose Badge

Strany 12 - Configuration

Caution !Store the Dose Badge in Stop Mode and ONLYin a fully charged state. To ensure optimumoperation, the Dose Badge should be left ontrickle charg

Strany 13

4. OPERATIONPractically all procedures for measuring noise dose require that themeasuring device be calibrated both before and after the measurement !

Strany 14

A Calibration ok message is displayed when calibration hasbeen completed satisfactorily.If the calibration cannot be completed sucessfully, the Reader

Strany 15 - 3.1 Set the Clock

A Resetting ok message is displayed when reset has beencompleted satisfactorily.At the same time as each Dose Badge is Reset, the configurationcurrent

Strany 16 - 3.3 Configure Criterion Time

The infrared window should face the front or side as shown to easestarting and stopping measurement sessions using the infrared link fromthe Reader.4.

Strany 17

Follow the starting progress on the Reader display.A red indicator will be lit in the badge window for up to twoseconds to show that the message has b

Strany 18 - 3.4 Configure Criterion Level

4. Handling the Dose Badge will now have no affect on themeasurements so it can be removed from the wearer.Most noise dose measurement procedures reco

Strany 19

The Reader downloads the measurements before it downloadsTime History. If there are errors during the download of TimeHistory data, the Reader will st

Strany 20 - 4.1 Automatic Calibration

6. Download data from the other Dose Badges used for the job inthe same way.4.8 Review Stored MeasurementsOnce the measurements have been downloaded f

Strany 21 - 4.2 Reset the Dose Badge

4.9 Download Measurements to a PCMeasurements that have been stored in the Reader can be downloadedto a PC using the dBLink software. The dBLink softw

Strany 22 - 4.3 Wearing the Dose Badge

Reader via the cable supplied. This software also supports USBconnections where an RS232 port is not available.PLEASE NOTE that when a USB connector i

Strany 23 - 4.4 Start Measurement

ContentsChapter ... Page1. INTRODUCTION . . . . . . . . . . . . . . . . . . . 51.1 The System ... 61.2 Layout &

Strany 24 - 4.5 Stop Measurement

When dBLink software is running, it will automatically detect theReader Unit and allow measurements to be downloaded.If the connection is not automati

Strany 25 - Dose Reader Unit

5. ROUTINE CARE5.1 Annual VerificationIt is recommended that the performance of all Dose Badges and ReaderUnits are verified and their batteries chang

Strany 26 - 4.7 Print Stored Measurements

The Reader shows the battery level as an icon in the top righthand corner of the display. As the battery level falls the icon will empty.Batteries for

Strany 27 - Operation

6. TROUBLESHOOTINGThe following information is designed to solve common problems foundwhen using the Dose Badge system. If the problem cannot be solve

Strany 28

6.2 Dose Badge Troubleshooting6.3 Charger TroubleshootingTroubleshooting

Strany 29

Troubleshooting

Strany 30

7. TECHNICAL INFORMATION7.1 SpecificationThe CEL-310 Dose Badge Personal Noise Dose Dosimeter and theCEL-315 Reader Unit have been designed to meet th

Strany 31 - 5.3 Storage

Technical Information

Strany 32 - Routine Care

7.2 Ordering InformationDose Badge measurement Kits can be ordered with standard numbers ofDose Badges.Number of Dose Badges Standard Measurement Kit2

Strany 33 - 6. TROUBLESHOOTING

8. SERVICE & WARRANTYTo ensure its conformity with the specification, this equipment isthoroughly inspected and it’s accuracy verified prior to di

Strany 34 - 6.3 Charger Troubleshooting

Chapter ... Page5. ROUTINECARE ...315.1 Annual Verification ...315.2 Cleaning ...315

Strany 35 - Troubleshooting

period of this equipment. At the end of the first warranty period, it isrecommended that the equipment be returned to the Service andVerification Depa

Strany 36 - 7.1 Specification

APPENDIX 1 - GLOSSARYThis Appendix gives a glossary of the acoustic terminology used in thishandbook and in the Dose Badge Personal Noise Dosimeter an

Strany 37 - Technical Information

Appendix 1Hours) or Pa2s (Pascal Squared Seconds).In the Dose Badge system, Pa2h are used.For example, a noise level of 94 dBfor 4 hours would be 1 Pa

Strany 38 - 7.2 Ordering Information

Appendix 1Assuming the noise level for the rest of the8 hour reference period is “quiet”, the LEP,dwill be:Lower than the Leqwhen themeasurement durat

Strany 39 - 8. SERVICE & WARRANTY

Appendix 1The laboratory and field grade for soundlevel meters and personal noise dosimetersThe general field grade for sound levelmeters and personal

Strany 40 - CASELLA USA

Appendix 1001 of 01017:05:28 20/06/0217:05:28 20/06/0217:05:28 20/06/02Serial No.Meas. timeMeas. DateRun TimeNNNNNN16:5506/02/0207:45:12Measurementsus

Strany 41 - Appendix 1

1. INTRODUCTIONThe CEL-310 Dose Badge is a unique solution to problems associatedwith the measurement of personal noise exposure. It is controlled and

Strany 42

The system is delivered with the Dose Badge configured foreither ISO or OSHA measurement. The user can change the configurationby means of the Reader

Strany 43

When a measurement is downloaded from the Reader to theDose Badge Software, the configuration of the Dose Badge is shown.PLEASE NOTE that the Micropho

Strany 44

An infrared port beside the acoustic calibrator in the top panel, isused to communicate with the Dose Badge. Ensure that this window isclean and free

Strany 45

2. PREPARATIONIt is recommended that the instructions and information contained in thishandbook be read before operating the equipment.2.1 Unpack the