CASELLA CEL CEL-490 Uživatelský manuál Strana 42

  • Stažení
  • Přidat do mých příruček
  • Tisk
  • Strana
    / 78
  • Tabulka s obsahem
  • KNIHY
  • Hodnocené. / 5. Na základě hodnocení zákazníků
Zobrazit stránku 41
CEL-63x Environmental Noise Meter
42 Casella CEL
The printed report includes the optional user logo; the run start date and
time and duration; and the instrument setup and calibration data. This is
followed by the cumulative or period broadband and narrowband results
and then by the Location, Operator and Signature fields which can be filled
in by the user.
NOTE
The functions printed in the report (and displayed on the results screens) are those that
are defined in the current measurement setup. The printing of octave and 1/3-octave
results can use considerable amounts of printer paper, especially if a number of
different periods are printed.
Up to five functions are printed for each set of frequency bands, so if Octave Leq +
Max is enabled then selecting more than three Ln functions in the current
measurement setup will require two sets of frequency bands to be printed, using more
paper. Disabling both Octave options above just prints a broadband set of results.
Print Screens
This mode captures the current result screen and prints a graphical copy of this
screen to the printer.
While displaying the Memory results screen, select the View icon (instead of the
Printer icon) and use the keys to select each screen of the current runs results.
Press the Run/Stop key when you want to print the current screen.
This printing format is particularly useful when displaying octave or 1/3-octave
results when the graphical bar graph of the spectrum can be printed for any of
the selected narrow band functions.
Any mix of print reports and result screens can be printed.
Printer set-up
It is necessary to set up the printer (-CMC73) before using it for the first time.
NOTE
Refer to the User and Programming manuals for the -CMC73 on the CD which is
supplied with the printer for details on setting the dip switches and the RS232 protocol.
The -CMC73 printer dip switches SW1, SW2, SW3 and SW4 should ALL be in
the OFF position, which selects the following options:
wide paper roll (78 mm)
ESC/POS mode
Hardware handshaking
Normal operation
The -CMC73 uses the following RS232 settings when used with a CEL-63x:
115200 baud